14 PFB-BA-e-13226.7.1 Procedure when adjusting (non-verifiable models):Observe stable environmental conditions.A warming up time (see chapter 1; Technical Data) is required for stabilization.Ensure that there are no objects on the weighing plate.Balance display Operation In weighing mode press approx. 3 sec. until„UnLoAd“ appears.When display „LoAd“ appears, put the necessaryadjustment weight (see chap.1; Technical Data) on thecenter of the weighing plate.Adjustment will be carried out automatically after deadstop control. Wait for stability display„Pass“ appears, the adjustment process has beenfinished successfully. Take away adjustment weightWait until the balance is again in the weighing mode.