PLJ-N-BA-e-1333 23Data output adjustment log:The content of the data output is determined in menu P2 GLP. All parameters set to“YES“ will be issued.Example:P2 GLPP2.1 uSr |P2.2 PrJ |P2.3 Ptin | YESP2.4 PdAt | YESP2.5 PuS | YESP2.6 PPrJ | YESP2.7 PId | YESP2.8 PFrn | YES7.3 VerificationGeneral introduction:According to EU directive 90/384/EEC balances must be officially verified if they areused as follows (legally controlled area):a) For commercial transactions if the price of goods is determined by weighing.b) For the production of medicines in pharmacies as well as for analyses in themedical and pharmaceutical laboratory.c) For official purposes.d) For manufacturing final packages.In cases of doubt, please contact your local trade in standard.Verification instructionsAn EU type approval exists for balances described in their technical data asverifiable. If a balance is used where obligation to verify exists as described above, itmust officially verified and re-verified in regular intervals.Re-verification of a balance is carried out according to the respective nationalregulations. The validity for verification of balances in Germany is e.g. 2 years.The legal regulation of the country where the balance is used must be observed!After verification the balance is sealed at the indicated positions.Verification of the balance is invalid without the "seal".