Instructions for use ESTETICA E307 Safety checks - testing instructions | 7.2 Instructions for safety checks120 / 1347.2.5 Assessment and documentationNoteAll tests conducted must be documented comprehensively. The doc-uments must contain at least the following particulars:▶ Name of the test centre▶ Name of the test engineer▶ Name of the tested device (e. g. type, serial number)▶ Tests and measurements▶ Data, type and measuring results of the visual inspections▶ Data, type and measuring results of the measurements▶ Data, type and measuring results of functional tests▶ Measuring/test equipment including SN/test equipment number and cal-ibration period▶ Final evaluation▶ Name, date and signature of test engineerThere is a copy of a test report template at the end of chapter STK. KaVo re-commends the use of this template.NoteFollowing testing, repair or adjustment, it must be verified whether the MEequipment or ME system has been restored to the state that is required forthe intended usage before it is employed once again.NoteIf the safety of the tested ME equipment or ME system has not beenestablished, e.g. the tests have not been completed with positiveresults, the equipment or system must be marked accordingly andthe potential hazard emanating from the equipment or system mustbe communicated in writing to the RESPONSIBLE ORGANISATION(to the operator, as a rule). This action is not required if the cause ofthe malfunction could be determined and rectified. The defect mustbe recorded in the protocol.