Instructions for use EXPERTsurg LUX REF 1.008.35007 Reprocessing steps in accordance with DIN EN ISO 17664 | 7.2 Disinfection7.1.2 Machine cleaningCAUTIONDamage by liquids.Faults on electrical components.▶ Do not machine-clean the device.The following parts of the unit are released for machine-based cleaning:▪ Handpiece tray▪ Motor and motor cableNotePlease comply with the corresponding Instructions for use when you re-process themotor and motor cable.The re-processing steps for the handpiece tray are as follows:KaVo recommends washer disinfectors in compliance with ISO 15883-1, which areoperated using alkaline cleaning agents.▶ For program settings as well as cleansers and disinfectants to be used, pleaserefer to the Instructions for Use of the thermodisinfector.7.2 DisinfectionNoteAfter each treatment of a patient, the surfaces near the patient that may have beencontaminated by contact or aerosol need to be disinfected. All disinfection meas-ures need to be carried out by wipe disinfection.NotePlease comply with the corresponding Instructions for use when you re-process themotor and motor cable.7.2.1 Manual disinfection▶ Use a soft disposable cloth and an approved disinfectant for disinfection by wipingdown all visible surfaces of the unit, bottle holder, foot control surfaces, and con-necting cables. Make sure that all surfaces are wetted.▶ Let the disinfectant act for the prescribed time.▶ Dry the surfaces.Permissible disinfectants (application range in accordance with the available manufac-turer's instructions and national guidelines. Please note material safety data sheets.)KaVo recommends the following products based on the compatibility of the materials.The microbiological efficacy must be confirmed by the disinfectant manufacturer.▪ FD 322 Dürr▪ Microcide AF Liquid (Schülke & Mayr)▪ CaviCide made by Metrex7.2.2 Automated disinfectionThe following parts of the unit are released for machine-based disinfection:57 / 72