Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.01007 Processing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing18 / 26WARNINGSharp tool in the medical device.Injury hazard from sharp and/or pointed tool.▶ Remove the tool.NOTICENever disinfect the handpiece with chlorine-containing products.Functional damage and property damage.▶ Only disinfect in a washer disinfector or manually.NOTICENever process this medical device in an ultrasonic device.Functional damage and property damage.▶ Clean manually or in a washer disinfector only.7.3.1 Automated internal and external cleaning andinternal and external disinfectionKaVo recommends thermodisinfectors in compliance with ISO 15883-1, whichare operated using alkaline cleaning agents having a maximum pH value of 10.▶ For program settings as well as cleansers and disinfectants to be used,please refer to the Instructions for Use of the washer disinfector.7.3.2 Automated DryingThe drying procedure is usually part of the cleaning program of the washer dis-infector.NotePlease comply with the instructions for use of the washer disinfector.▶ In order to prevent impairment of the KaVo medical device, make sure thatthe inside and outside of the device is dry after the end of the cycle.▶ Immediately after drying, lubricate the KaVo medical device with careagents from the KaVo care system.7.4 Care products and systems - ServicingWARNINGSharp tool in the medical device.Injury hazard from sharp and/or pointed tool.▶ Remove the tool.CAUTIONImproper service and care.Risk of injury.▶ Perform regular proper care and servicing.