Instructions for use EXPERTtorque E680 - 1.006.4700EXPERTtorque Mini E677 - 1.006.01007 Processing steps in accordance with ISO 17664 | 7.4 Care products and systems - ServicingNoteKaVo only guarantees that its products will function properly if the care prod-ucts listed as accessories are used, since these products have been tested forproper use on our products.7.4.1 Servicing with KaVo SprayKaVo recommends servicing the product as part of the processing after eachuse, i.e. after each cleaning, disinfection, and before each sterilization.▶ Remove the tool from the medical device.▶ Cover the medical device with the KaVo Cleanpac bag, and place it on thecorresponding care adapter.▶ Press the spray key once for 1-2 seconds.Servicing the chuckKaVo recommends servicing the chucking system once weekly.▶ Remove the tool from the medical device.▶ Position the tip of the spray nipple in the opening, and apply the spray.▶ Press the spray key once for 1-2 seconds.7.4.2 Servicing with KaVo QUATTROcare PLUSServicing and cleaning device with expansion pressure for internal cleaning ofinorganic residues and optimum servicing.KaVo recommends servicing the product as part of the processing after eachuse, i.e. after each cleaning, disinfection, and before each sterilization.▶ Remove the tool from the medical device.▶ Service the product in the QUATTROcare PLUS.See also:2 Instructions for use KaVo QUATTROcare PLUSServicing the chuckKaVo recommends servicing the chuck system once a week using the chuckservicing program integrated in the device.NoteHandpieces must be taken off the service couplings before the chuck servicecan be started and run.▶ Close the front door and press the chuck service button for at least threeseconds until the spray canister control LED flashes three times consecu-tively.ð The device is in chuck service mode.19 / 26