5 Imaging programs5.3 3D programsNOTICE! These programs are available only for 3D devices.Always select the smallest feasible FOV size, resolution and imaging parameters for the3D image in order to follow the ALARA (As Low As Reasonably Achievable) principle.NOTICE! It is always up to the dental professional to select the appropriate FOV,resolution and imaging parameters.5.3.1 FOV SizesThe indicated FOV sizes are Height x Diameter of the volume in centimeters. Forexample FOV 5 x 5 corresponds to a cylindrical volume with diameter of 5 cm and heightof 5 cm.NOTICE! Fine adjustment of the FOV location and FOV height can be done usingthe scout images. The location of the FOV is determined by the selected ROI.FOV 5 x 5Optimized for single site implants or localized diagnostics, for example, 3rdmolar extractions, impacted teeth, periodontal cases, root fractures, singleTMJ analysis, endodontics and for pediatric imaging.Available resolutions:LDT (Low Dose Technology™) resolutionStandard resolutionHigh resolutionResolution optimized for endodontic imaging:Endo resolutionFOV 6 x 9Optimized for multiple implant placement using surgical guides, imaging thewhole dental arc of one jaw, 3rd molar visualization, pathology (bilateralanalysis) and periodontal cases.Available resolutions:LDT (Low Dose Technology™) resolutionStandard resolutionHigh resolutionORTHOPANTOMOGRAPH™ OP 3D 43D519207, 19Copyright © by Palodex Group Oy. All rights reserved.See PDM system to determine the status of this document. Printed out 2019-10-07 09:30:50ApprovedReviewed Suuronen Esa 2019-09-19 12:21Approved Lassinen Juuso 2019-10-02 14:37