NXR-800H124ADJUSTMENTItem ConditionMeasurement AdjustmentSpecifications / RemarksTest-equipment Unit Terminal Unit Parts Method24. DQTDeviation1) Channel: 3 (Center)Wide/Narrow: Analog WideConnect the deviation meterto the TX ANT end via theATT.Deviation meter settingHPF: OFFLPF: 3kHzDe-emp: OFFDetector: Peak holdDeviationmeterRear TX ANT PCADJ±0.75kHz ±0.05kHzConnect the deviation meterto the TX ANT end via theATT.2) Channel: 3 (Center)Wide/Narrow: Analog Narrow±0.35kHz ±0.05kHz25. CW IDDeviation1) Channel: 3 (Center)Wide/Narrow: Analog WideConnect the deviation meterto the TX ANT end via theATT.Deviation meter settingHPF: OFFLPF: 15kHzDe-emp: OFFDetector: +peak, –peakDeviationmeterRear TX ANT PCADJ±2.00kHz ±0.05kHzConnect the deviation meterto the TX ANT end via theATT.2) Channel: 3 (Center)Wide/Narrow: Analog Narrow±1.00kHz ±0.05kHz26. Test ToneDeviation1) Channel: 3 (Center)Wide/Narrow: Analog WideConnect the deviation meterto the TX ANT end via the ATT.Deviation meter settingHPF: OFF, LPF: 15kHzDe-emp: OFFDetector: (p-p)/2DeviationmeterRear TX ANT PCADJ±3.00kHz ±0.05kHzConnect the deviation meterto the TX ANT end via theATT.2) Channel: 3 (Center)Wide/Narrow: Analog Narrow±1.50kHz ±0.05kHz27. DTMFDeviation1) Channel: 3 (Center)Wide/Narrow: Analog WideConnect the deviation meterto the TX ANT end via the ATT.Deviation meter settingHPF: OFF, LPF: 15kHzDe-emp: OFFDetector: (p-p)/2DeviationmeterRear TX ANT PCADJ±2.85kHz ±0.05kHzConnect the deviation meterto the TX ANT end via theATT.2) Channel: 3 (Center)Wide/Narrow: Analog Narrow±1.50kHz ±0.05kHz28. CourtesyToneDeviation1) Channel: 3 (Center)Wide/Narrow: Analog WideConnect the deviation meterto the TX ANT end via the ATT.Deviation meter settingHPF: OFF, LPF: 15kHzDe-emp: OFFDetector: (p-p)/2DeviationmeterRear TX ANT PCADJ±1.00kHz ±0.05kHzConnect the deviation meterto the TX ANT end via theATT.2) Channel: 3 (Center)Wide/Narrow: Analog Narrow±0.50kHz ±0.05kHz