1- 16 85058B/EGeneral InformationIncoming InspectionIncoming InspectionVerify that the shipment is complete by referring to Figure 6-1 on page 6-5 or Figure 6-2 onpage 6-7.Check for damage. The foam-lined storage case provides protection during shipping.If the case or any device appears damaged, or if the shipment is incomplete, contact Keysight.See “Contacting Keysight” on page 5-5. Keysight will arrange for repair or replacement ofincomplete or damaged shipments without waiting for a settlement from the transportationcompany.When you send the kit or device to Keysight, include a service tag (found near the end of thismanual) with the following information:• your company name and address• the name of a technical contact person within your company, and the person's completephone number• the model number and serial number of the kit• the part number and serial number of the device• the type of service required• a detailed description of the problemOSL 35 to60 GHz0.052 0.062OSL 60 to67 GHz0.055 0.055a. Expanded Math has no effect with the 85058E data-based model or the85058EP polynomial model.b. Refer to “Two Models for Defining Calibration Standards” on page 1-2.c. Using the 85058E Data-Based.ckt (PNA) calibration definitions file.d. Using calibration definitions file 85058EP Polynomial.ckt.Table 1-8 (Continued) 85058E (Economy) Residual System Calibration Error SpecificationsParameter CalMethodaFrequencyRangeResidualsPNA Data-BasedModelb,cPNA PolynomialModel b,d