2- 4 85058B/ESpecificationsElectrical Specificationsits pin depth. The electrical specifications for each device in the kit take into account the effectof pin depth on the device’s performance. Table 2-2 on page 2-4 lists the typical pin depths andmeasurement uncertainties, and provides observed pin depth limits for the devices in the kit. Ifthe pin depth of a device does not measure within the observed pin depth limits, it may be anindication that the device fails to meet electrical specifications. Refer to Figure 2-1 for a visualrepresentation of proper pin depth (slightly recessed).Table 2-2 Pin Depth LimitDevice Typical Pin Depth MeasurementUncertaintyaa. Approximately +2 sigma to -2 sigma of gage uncertainty based on studies done at the factoryaccording to recommended procedures.Opens -0.005 to -0.015 mm-0.0002 to -0.00059 in+0.0030 to -0.0030 mm+0.00012 to -0.00012 in-0.0020 to -0.0180 mm-0.00008 to -0.00071 inShorts 0 to -0.0127 mm0 to -0.0005 in+0.0015 to -0.0015 mm+0.00006 to -0.00006 in+0.0015 to -0.0142 mm+0.00006 to -0.00056 inFixed LoadsandTerminations-0.0025 to -0.0203 mm-0.0001 to -0.0008 in+0.0030 to -0.0030 mm+0.00012 to -0.00012 in+0.0005 to -0.0233 mm+0.00002 to -0.00092 inAdapters 0 to -0.0381 mm0 to -0.0015 in+0.0030 to -0.0030 mm+0.00012 to -0.00012 in+0.0030 to -0.0411 mm+0.00012 to -0.00162 inElectrical SpecificationsThe electrical specifications in Table 2-3 through Table 2-4 apply to the devices in yourcalibration kit when connected with a Keysight precision interface.Observed Pin DepthLimits bb. Observed pin depth limits are the range of observation limits seen on the gage reading due tomeasurement uncertainty. The depth could still be within specifications.