3-18 Keysight B1505A Configuration and Connection GuideN1265A Connection GuideOutput ConnectionHC/MC/DHCSMU via SMU 3 terminals should be connected as follows.• SMU 3 High Force and Sense to DUT high• SMU 3 Low Force and Sense to DUT low• GNDU Force and Sense to DUT lowThe GNDU signals will appear at the Selector Output Low terminals as shown inTable 3-7.3. Set the DUT on the socket.4. Close the fixture cover and perform measurement.Universal Socket Module (N1265A-011)This is a blank module, kind of a do-it-yourself kit for supporting variety of packageddevices. This module can be used by mounting your desired socket or packaged device andmaking connections same as the N1265A-010 Inline Package Socket Module.To use this module, see the following simple instruction. For the component locations anddimensions, see Figure 3-8. Also see "Inline Package Socket module (N1265A-010)" onpage 3-17 to perform measurement.Figure 3-8 Universal Socket Module1. Prepare the following parts.• blank board suitable for mounting the socket or packaged device• screw M3, 4 ea., for fixing the blank board on the socket module• wire, adequate length and quantity, for making connections• hexlobe (torx type) screwdriver T-10• socket, if you use, and packaged device under test (DUT)2. Cut the blank board in 90 mm 81 mm square.3. Make four screw holes on the board. The holes should be 6 mm inside from the edge.4. Fix the board to the blank module.5. Remove the cover bottom of the blank module.165 mm165 mm129 mm160 mm2 1345678Collector / Drain Emier / Source Base / GateN1265A Opt 011Force Sense Force Sense Force SenseUniversal Socket Module1 25 64391 mm (Nut pitch: 78 mm)(Nut pitch: 69 mm)82 mm