7-40 Keysight B1505A Configuration and Connection GuideConnection and Ordering ExamplesConfiguration Examples for Lateral Device Measurement with Wafer Prober3 kV, 500 A, Capacitance Measurement for On-WaferLateral Device• Device Type: Lateral MOSFET (4 terminals)• Key Specification• Max voltage/current: 3 kV/500 A• Capacitance measurement up to 3 kV DC bias• Module Selector for multi-parameter automated measurement functionFigure 7-21 Connection example for CV measurementNOTE The connection diagram in Figure 7-16 shows Cds measurement.NOTE For IV measurement connection diagram, please refer to Figure 7-20.NOTE Prober vendor is responsible for cabling inside the shielding box.Table 7-19 Ordering exampleModel/Option Quantity DescriptionB1505A 1 Power Device Analyzer/Curve Tracer mainframeB1513C-FG 1 High Voltage Source Monitor Unit, 3000 V/4 mA (Pulsed & DC)B1514A-FG 4 Medium Current Source Monitor Unit, 1 A/30 V(Pulsed), 100 mA/30 V(DC)B1520A-FG 1 Multi Frequency Capacitance Measurement Unit Module6316494J Interlock (Included in B1505A)16493G-001 Digital I/O cable (Included in N1265A)DGSHVSMUGNDUMCSMUB1505AMCSMUMCSMUFSFSFSHVSMUGNDUGate FGate SUHC InputV ControlI ControlFor GateFor UHCFor UHCFor DrainN1265AMCSMUFSChuckFor ChuckN1260ABias-TTo change connecƟonmanuallyHcHpLcLpHV-F AC-HAC-LAC-GuardMFCMU N1300A (Included in the MFCMU)N1254A-51816494T (Included in the HVSMU)16494T (Included in the HVSMU)16493L (Included in B1505A)16494A(Included in the MCSMU)HCSMU Non-Kelvin adapter16493S-011N1254A-104 TRX(J) to BNC(P) adapter16493U-00116493U-001