Keysight B1505A User’s Guide, Edition 12 5-43Measurement ExamplesTemperature Controlled by ThermoStream10. Define the test setup execution order in the quick test.Click the Quick Test tab on the main screen. The tab is in the leftmost column onthe screen.The Quick Test screen lists the test setups created in steps 2 to 9.If they are not listed according to the following order, change the order using theUp/Down buttons.a. ThermoStream Initialize (for initialization)b. ThermoStream DefineParameters (for setting up the operation parameters)c. ThermoStream ControlTemperature (for controlling the airflow temperature)d. ThermoStream WaitTemperature or N1265A WaitTemperature (for waiting)e. A device test setupf. Repeating of steps c, d, and eg. ThermoStream ControlTemperature (for ordinary temperature)h. ThermoStream FlowOff (for stopping the airflow)Performing Temperature Control and Measurement1. Set your DUT into the N1271A-002 or -005 enclosure, and make theconnections.2. Adjust the head position of the ThermoStream for the top cover surface of theN1271A-002 or -005 enclosure, and then dock the ThermoStream with theenclosure.3. Click the Single button to start Quick Test. This opens the Data Displaywindow.A temperature at the measurement is added to the Device ID.WARNING Activating the Thermostream may cause low or high temperature of -50 °C to+220 °C on the enclosure front panel, the cover, and the left and rightprotection panels. To prevent yourself from getting injured, confirm that theyare in the safe temperature range before touching the enclosure.