What's New 1Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 49ChangedCommandsDiscontinuedCommandsCommand Differences:CHANnel:PROBe (seepage 354)The probe attenuation factor can now be set from 0.001:1 to10000:1.:FUNCtion[:FFT]:VTYPe(see page 482)With the FFTPhase operation, you can select vertical units inDEGRees or RADians.:FUNCtion:OPERation (seepage 490)The FFTPhase operation is added.:FUNCtion:SOURce1 (seepage 498)Reference waveforms can now be specified as source 1.:FUNCtion:SOURce2 (seepage 500)Reference waveforms can now be specified as source 2.:POWer:HARMonics:LINE (seepage 739)The AUTO frequency option is added. This option automaticallydetermines the frequency of the Current waveform.:SBUS:LIN:STANdard (seepage 980)The LIN13NLC option is added for selecting LIN 1.3 (no lengthcontrol).:SBUS:MODE (seepage 857)The CXPI mode is now available with the CXPI serial decode andtriggering license.:TIMebase:REFerence (seepage 1287)The CUSTom option is added for placing the time referencelocation at a percent of the graticule width.Discontinued Command Current Command Equivalent Comments:POWer:CLResponse:VIEW none Because the plot now appearsin its own dialog box andcontains both gain and phaseplots, there is no longer a needto select one or the other.:POWer:CLResponse:YMAXimumnone Specifying the plot's initialvertical scale maximum valueis no longer necessary becausethe plot now appears in its owndialog box and is autoscaled.:POWer:CLResponse:YMINimumnone Specifying the plot's initialvertical scale minimum value isno longer necessary becausethe plot now appears in its owndialog box and is autoscaled.:POWer:PSRR:RMAXimum none Specifying the plot's initialvertical scale value is no longernecessary because the plotnow appears in its own dialogbox and is autoscaled.