4 Multimeter Setup Options136 Keysight U1273A/U1273AX User’s GuideTo change the temperature unit:1 Press for more than 1 second to enter the multimeter’s setup menu.2 Browse to Menu 2 > T-UNIT. Press and hold for more than 1 second tounlock this setting, then press edit the value.3 Use the arrow keys to change the temperature unit.4 Press to save your changes (or press to discard your changes).5 Press and hold until the multimeter restarts to return to normal operation.Changing the % scale rangeThis setting is used with % scale current measurements (see page 97). Themultimeter converts DC current measurements to a percentage scale readout of0% to 100% based on the selected range in this menu. For example, a 25%readout represents a DC current of 8 mA on the 4-20 mA % scale, or a DC currentof 5 mA on the 0-20 mA % scale.To change the % scale range:1 Press for more than 1 second to enter the multimeter’s setup menu.2 Browse to Menu 2 > mA SCALE, and press edit the value.3 Use the arrow keys to change the % scale range. Select OFF to disable the %scale readout.4 Press to save your changes (or press to discard your changes).Parameter Range Default settingT-UNIT °C, °F/°C, °C/°F, or °F °CSetupShiftViewEscHz % msLogHz % msLogShiftViewEscSetupParameter Range Default settingmA SCALE 4-20 mA, 0-20 mA, or OFF 4-20 mASetupHz % msLogHz % msLogShiftViewEsc