Chapter 3: Getting StartedProduct Service 1-800-522-7658184. Remove the cover over the electrical junction box by removing the twoscrews.5. Make sure electrical power is turned off before making any connections.6. On 115V models, connect the customer-supplied power wires to the black andwhite wires, using wire connectors. On 230V models, attach customer-supplied power wires to the brown and blue wires, using wire connectors.Customer-supplied wires must be clamped to the hole where they passthrough the rear of the junction box, if prescribed by local codes. If it isnecessary to use a bushing with an integral clamp at the inlet to the junctionbox, remove the plastic bushing supplied with the washer before installing thenew customer supplied bushing.7. Connect the customer-supplied ground wire to the green screw.8. Replace the cover on the electrical junction box.9. Replace the lower panel and door panel assembly on the washer.Controller Set-UpThe washer control system has been set at the factory to agree with theconfiguration of the washer as it was shipped. The control may be customized ormodified if certain optional accessories are added.