Chapter 6: Modifying Your WasherProduct Service 1-800-522-765842Installing Small SpindlesTen small spindles are included with each Spindle Rack. Since the large volumeof water which flows through the standard spindles may be great enough to forcethe glassware off the spindle, small spindles should be used when small pieces ofglassware will be washed.To install small spindles:1. Remove the standard spindle by inserting the tool supplied with the SpindleRack or a screwdriver into the slot on the top and turning counterclockwise.2. Replace the standard spindle with a small spindle.Installing the Optional Cool Drain Water KitAn optional Cool Drain Water Kit is available to reduce the water temperature ofthe drain water in installations where hot drain water is not permitted. It may beinstalled on Undercounter or Freestanding models. A supply of cold water mustbe available. It is recommended that the cold water temperature is colder than54°F (12°C). Warmer water temperature will cause the drain temperature to behigher. The cold water supply must be capable of a flow rate of 3.5 GPM (13.2LPM) and a pressure of 20-120 PSI (138-827 kPa). Hot drain water will be mixedwith cold water during drain operations. To install the optional Cool Drain WaterKit, refer to the figure below and the following instructions.To install the optional Cool Drain Water Kit:1. Disconnect power to the washer.2. Remove the back panel of the washer if it is a Freestanding model or the coverpanel of an Undercounter model.3. Attach the water valve assembly to the rear flange of the base using the screwsprovided.4. Connect the wires coming from the water valve assembly to the matingconnector in the wiring harness of the washer.5. Connect the cold water supply to the water valve assembly. Do not allow thetube to kink when the washer is placed into its operational location. Ifnecessary, replace the straight fitting on the cool water valve with the suppliedelbow.6. Replace the back panel if it is a Freestanding model or the cover panel if it isan Undercounter model.7. Attach the hose from the valve to a Twin Inlet High Flow Air-gap suitable fordishwasher applications. This may be obtained from a plumbing supply store.8. Attach the drain hose from the Glassware Washer to the other inlet of the Air-gap.9. Attach the Air-gap to the building drain.10. Turn on the electrical power.