– 7 –© LEYBOLD VAKUUM GMBH GA 05.212/10 - 09/00DESCRIPTION 11.2 Technical dataSupply voltage ranges .............................. 100-140 V AC,-10%/+16%Can be switched-over to .................................. 200-240 V AC, ±10%Supply frequency ................................................................ 50/60 HzLine fuse (F1 internal) .................................................... T 1A / 250 VRated speed of the TURBOVAC ................... 72,000 RPM = 1200 HzBriefly when accelerating .................................... up to 64 VA (45 W)Continuously in operation (rated speed) ...................... 35 VA (18 W)Idle operation (no-load) .................................................... 6 VA (3 W)Voltage ............................................................................... 3 x 150 VSupply frequency ................................................................. 1200 HzFrequency range of the output voltage .......................... 220-1250 Hzin operation ...................................................................... 0°C - 40°Cstorage ...................................................................... -40°C to +85°CSwitching voltage ........................................... £ 250 V AC, £ 30 V DCSwitching current ................................................ £ 6 A AC, £ 5 V DCSwitching power for DC ......................................................... £ 150 WSwitching power for AC ...................................................... £ 1500 VAThe units have degree of protection IP20 in accordance withEN 60 529. An increased degree of protection, e.g. IP54 can only berealized by mounting the units in other housings.TURBOTRONIK NT 10 ............................................................. 1.5 kgTURBOTRONIK NT 12 ............................................................. 1.5 kgTURBOTRONIK NT 13 ............................................................. 1.5 kgTURBOTRONIK NT 10 ................................................. 1/4 19”, 3 HETURBOTRONIK NT 13 ................................................. 1/4 19”, 3 HE1.3 Scope of supplyThe connecting cable to the TURBOVAC is not included in thescope of supply!1.3.1 TURBOTRONIK NT 10TURBOTRONIK NT 10, miniature fuses, mating connector for thePhönix connector (control terminal strip X10), attachable feet fordesktop operation, 4 adhesive feet, four M 3x8 screws, the powercable is permanently attached.1.3.2 TURBOTRONIK NT 12TURBOTRONIK NT 12, miniature fuse.1.3.3 TURBOTRONIK NT 13TURBOTRONIK NT 13, miniature fuses, 4 screws for rackinstallation.Power consumptionOutputsAmbient temperatureMax. load capability ofthe relay contactsWeigtDimensions andhousing dimensions(refer to Fig. 4)Degree of protectionSupply