878717. TECHNICAL DATAMain inputInput signals and measuring ranges Table 191) The intrinsic error is related to measuring range 200…1767 °C(392...3212,6 °F)Sensor type Standard Range Sym-bolPt100 EN60751+A2:1997-200...850 °C -328...1562 °F pt1Pt1000 -200...850 °C -328...1562 °F pt10Fe-CuNi (J)EN 60584-1:1997-100...1200 °C -148...2192 °F t-,Cu-CuNi (T) -100...400 °C -148...752 °F t-tNiCr-NiAl (K) -100...1372 °C -148...2501,6 °F t-kPtRh10-Pt (S) 0...1767 °C 32...3212,6 °F t-sPtRh13-Pt (R) 0...1767 °C 32...3212,6 °F t-rPtRh30-PtRh6 (B) 0...1767 °C 1) 32...3212,6 °F 1) t-bNiCr-CuNi (E) -100...1000 °C -148...1832 °F t-eNiCrSi-NiSi (N) -100...1300 °C -148...2372 °F t-nChromel – Kopel (L) GOST R 8.585-2001-100...800 °C -148...1472 °F t-llinear, current (I) 0...20 mA 0...20 mA 0-20linear, current (I) 4...20 mA 4...20 mA 4-20linear, voltage (U) 0...5 V 0...5 V 0-5linear, voltage(U) 0...10 V 0...10 V 0-10