Operation Manual touchTymp MI 24 and MI 34 Version8107420 Rev. 6 13 03/09/2018In order to maintain a high level of safety and to ensure thedevice works properly, it is necessary to have the deviceand its power supply checked according to the medicalelectrical safety standard IEC 60601-1 by a qualifiedservice technician at least once a year. For moreinformation see section 3.2.The use of non-calibrated devices can lead to incorrect testresults and is not advisable.Prevent cable breakage: cables must not be bent or buckled.2.7 Device ControlThe user of the device should perform a subjective device check once a weekaccording ISO 8253-1.For annual calibration see section 2.5 and 3.2.2.8 Electromagnetic Compatibility (EMC)Electrostatic discharge (ESD) according to IEC 61000-4-2.Use the device only in an electrostatic controlledenvironment.To avoid the risk of electric shock, this equipment mustonly be connected to supply mains with protective earth.The device fulfills the relevant EMC requirements. Avoidunnecessary exposure to electromagnetic fields, e.g. frommobile phones etc. If the device is used adjacent to otherequipment it must be observed that no mutual disturbanceappears.Please also refer to EMC consideration in section 6.5.CAUTIONWARNING