CKW-55 CHECKWEIGHER EN-17Press the TARGET button to initiate check weigh entry sequence. UNDER flashes momentarilyto indicate the first parameter to be entered. The last Yellow and the first Green Bar Graphsegments are lit during entry.The display flashes the last Minimum range value entered. Use the numeric keypad to entera new Minimum range value.Note: -NO- is displayed if the entered value is out of range or is not compatible with thegraduation settings, (refer to Section 3.6.1).Press TARGET to accept the entry. The display will freeze momentarily to confirm the entry.OVER flashes momentarily to indicate the parameter to be entered next. The last Green andthe first Red Bar graph segments are lit during entry.The display flashes the last maximum range value entered. Use the numeric keypad to entera new Maximum range value or press C to clear. .Press TARGET to accept the entry. The display will freeze momentarily to confirm the entry.END appears momentarily to confirm end of checkweigh entry sequence. The Indicatorreturns to active checkweighing mode.NOTES: 1.Values entered can be from 1d to Full Scale capacity.2.Minimum range value cannot be greater than maximum range value unless bothvalues are negative.3.For entry in the lb:oz units, the first decimal point entered is used as the lb ozseparator, (eg. an entry of 2.08.3 = 2 lb 8.3 oz. At least one digit must beentered for the lb location, eg., an entry of 0.15.9 = 0 lb 15.9 oz.Programming Accept Range by Numeric Key EntryIn the Mode menu, set Checkweigh to On and Set Point Input to Range (refer to Section 3.6.3).NOTE: The under and over set points are included in the Accept range.