EB SERIES EN-116. CALIBRATIONFor best results, calibrate the scale at regular intervals. Temperature changes,gravity variations, altitude changes and abuse are few reasons why a scale mayneed recalibration.When the scale is ideally positioned for operation, enter calibration and proceed asfollows:Long-press ZERO and COUNT at the same time to initiatecalibration. The required calibration weight flashes on thedisplay (ex. 3000g).At this time, a different calibration weight value can be selected by pressing .Available selections are 1/3, 2/3 and 100% of full capacity.Once the desired value is displayed, place the corresponding weight on the pan.The display flashes until the actual weight is registered and the calibration processends (the scale re-starts). At this time, remove the weight on the pan before thedisplay count-down sequence ends.NOTES:• Calibration weight is displayed in grams only.• To abort the calibration process, turn the scale off.• If needed, the factory default calibration value can be recovered. As thescale is turned on, press TARE and COUNT at the same time during the power-onsequence. The power-on sequence will re-cycle and the scale will reinitiate with thedefault calibration.