EN-10 FrontierTM Centrifuge FC5707Figure. 21 Figure. 22For adaption of data press the key "Store" (2) (refer to figures 21 and 22) for approx. 1 second. If the program isstored correctly, the word "StorE" appears in the display "RPM/RCF" (A-1). As a result, the word "programm"(M10) disappears.As soon as the key "Store" (2) is released, the word "programm xx" (M10) reappears – the (xx) stands for thechosen program location.If all program numbers are occupied you can take an old number that is not necessary anymore and just put in thenew parameters.3.6.2 Recall of stored programsTo recall stored programs press the key "Prog/Setup" (3) (refer to figure 23) while the lid is already closed. Insidethe display "Time/Prog" (A-4), "programm –"(M10) appears. The desired program number can be pre-selectedwith the adjusting dial (4).In the respective displays the stored values for that program will appear.If the wrong rotor is inside the centrifuge for the pre-selected program, in the display "RPM/RCF" (A-1), the word"rotor" (M3) flashes. At the same time the word "FALSE" and the stored rotor number "xx" (M4) will flashing byturns.Figure. 233.6.3 Leaving program modeTo leave the program mode just press the key "Prog/Setup" (3) (refer to figure 23). Then inside the display"Time/Prog" (A-4) the word "programm" (M10) appears.Set the display to "programm--" (M10) with the adjusting dial (4).324A-4M10A-1M3 M4