PX Series Balance EN-211. Weigh the calibrated sinker in the air with the balanceand the density determination kit.When the * (symbol of stability) appears, press theFunction / Mode button to confirm the weight of thecalibrated sinker in the air.2. Weigh the calibrated sinker submerged in the liquidwith the balance and the density determination kit.Lower the calibrated sinker down into the liquid until itis fully submerged (1 cm below the surface of theliquid).When the * (symbol of stability) appears, press theFunction / Mode button to confirm the weight of thecalibrated sinker. The density of the liquid is displayed.3. After completion of the test, press the Function / Modebutton to test a new sample.4.5.5 Measuring the Density of Porous Material Using OilPress and hold the Function / Mode button until Densityappears. Press Yes to enter the Density Determination.Item Settings:x Sample type: Solidx Auxiliary Liquid: Waterx Porous Material: On