7000 Series Indicators EN-27COUNtPC AUTO OPTONPC.OPtOFFmMOdEPERCENTONPErCNtOFFmMOdEDYNAMICAUTOSEMI AUTOMANUALdYNAmMOFF60LEVEL0mMOdECHECKWEIGHPCSWEIGHCHECKOFFmMOdEEND MODE3.6.4 Parts Counting OptimizationWhen the Parts Counting mode is turned ON, Parts Counting Optimization can be used toautomatically adjust the average piece weight (APW). Each time a quantity greater than 1x or lessthen 3x the previous quantity is placed on the scale, the APW is adjusted.OFF = disabledON = enabled3.6.5 Percent Weighing ModeSet the status.OFF = disabledON = enabled3.6.6 Dynamic Weighing ModeSet the status.OFF = disabledMANUAL = averaging and resetting are manually initiated by pressing theFUNCTION button.SEMI AUTO = averaging is automatically initiated when the display exceeds 5graduations; resetting is manually initiated by pressing the FUNCTIONbutton.AUTO = averaging is automatically initiated when the display exceeds 5graduations; resetting is automatically initiated when the display returnsto within 5 graduations of zero.If MANUAL, SEMI AUTO, or AUTO is selected, the current averaging time is displayed.Set the averaging time.Settings of 0 to 60 seconds are available.NOTE: Select 0 seconds to enable the Display Hold function. In this case, the first stable weight willbe held on the display.3.6.7 Check Weighing ModeSet the status.OFF = disabledWEIGH = enabled for checking items by weight.PCS = enabled for checking items by count.3.6.8 End ModeAdvance to the next menu or return to the top of the current menu.