EN-54 7000 Series IndicatorsPC003SAVE RECORD?P.tArE0.000 KGP.tArE1.000 KG*1.000 kg0.000 KG*1.000 kg1.000 KGStAtS kgN 1StAtS kgTOT 1.000StAtS kgSTD 0.000StAtS kgMIN 1.000StAtS kgMAX 1.000StAtS kgDIFF 0.000StAtS kgAVG 1.000StAtS kgCLEAR STATS?*1.000 kg0.000 KGIf desired, edit the displayed data value using the keypad. Then press the Yes button.After all data types have been viewed, press the Yes button to save the changes.4.11 Accumulation and StatisticsThe Accumulation feature enables manual or automatic totalizing of displayed values. Statistical datais stored in memory for review and printing.NOTES: ACCUMULATE must be set to MANUAL or AUTO in the SETUP menu.The Accumulation function is available in Weighing, Counting, Dynamic and Check Weighingmodes. Accumulation data is stored separately for each mode.To include the statistics data when printing, INFORMATION must be set to ON in the PRINTCONTENT menu.4.11.1 Accumulating Displayed ValuesWith ACCUMULATE set to MANUAL, place the item on the scale and press the FUNCTION button toaccumulate the displayed value.With ACCUMULATE set to AUTO, place the item on the scale. The displayed value is accumulatedautomatically.The accumulated value is displayed on the second line.4.11.2 Viewing Statistics DataTo view the statistics data, press the INFO button. The following statistics data will be displayedmomentarily: number of weighings, total, average, standard deviation, minimum, maximum, anddifference.4.11.3 Clearing Statistics DataTo clear the accumulation value and statistics data, press the CLR button while the statisticsinformation is being displayed. When the display shows CLEAR STATS?, press the Yes button toclear the statistics data or press the No button to keep the statistics data.