30■ CALIBRATIONThe CALIBRATION function in this unit measures LCD panel characteristics from low to high brightness valuesand internal monitor processing handles CALIBRATION.CALIBRATION does not rely on image quality settings since internal signals are used for a calibration.CALIBRATION in this unit is made at D65 color temperature and calibrations for other color temperatures areresults calculated from this value.■ Equipment required for calibration• Konica Minolta CA-210 display color analyzer• Konica Minolta CA-PU12 or CA-PU15 standard measurement probe• RS-232C cable (male to male, straight)• Connect the RS-232C terminal on this unit to the RS-232C terminal on the CA-210 display color analyzer using astraight cable.• Be sure to set the following calibration values on the CA-210 display color analyzer. The analyzer will not producea correct calibration unless it has been calibrated. For details on setup procedures, refer to the CA-210 User'sGuide.• Turn on this unit and perform adequate aging (about 1 hour) before starting calibration.• Turn off the lights in the room and make sure that no external light can enter the standard measurement probebefore starting the calibration. External light that enters the probe will prevent correct calibration of low brightnesscharacteristics.• LCD panel characteristics and instrument error in the display color analyzer may sometimes result in smalldifferences in values after calibration.In a fine tuning of the monitor, also set GAIN and BIAS for R, G and B in the COLOR TEMP. VAR mode.• Do not apply the probe to a WFM/VECTOR display on the screen.CA-210 side (Straight) BT-LH2550 SidePin number Signal Pin number Signal1 CD 1 N.C.2 RXD 2 TXD3 TXD 3 RXD4 DTR 4 DSR5 GND 5 GND6 DSR 6 DTR7 RTS 7 CTS8 CTS 8 RTS9 GND 9 N.C.Konica Minolta CA-210 displaycolor analyzerThis unitKonica Minolta CA-PU12 or CA-PU15standard measurement probeRS-232C cableCalibration W R G Bx 0.3091 0.6759 0.2029 0.1495y 0.3429 0.3126 0.691 0.0587Lv 195.3 47.56 135.2 13.21Main Menu (continued)