Maintenance and Performance Checks 5-72. Connect the Model 3152 output to the oscilloscope input.Use the 20 dB attenuator and set the oscilloscope inputimpedance to 50S.3. Set the oscilloscope and verify that the rise and fall timesare less than 6 ns. Verify that overshoot and ringing are lessthan 5% of amplitude.Sine Characteristics Specified Total Harmonic Distortion: <0.1% from 10.00Hz to100.0KHz (with a 4000 point waveform).Specified Harmonic Signals:<5MHz, <10Vp-p, -50dBc<5MHz, <16Vp-p, -45dBc<10MHz, <10Vp-p, -40dBc<10MHz, <16Vp-p, -35dBc<50MHz, <10Vp-p, -28dBc<50MHz, <16Vp-p, -23dBcEquipment: Krohn-Hite 6900 (or equivalent) Distortion Analyzer,RF Spectrum Analyzer, 50Ω feedthrough termination (or 20dB 50Ωattenuator if required).1. Connect the Model 3151 output to distortion analyzerinput.2. Set the distortion analyzer to distortion measurements, setthe Model 3151 frequency setting, and verify thedistortion reading as below.3. Configure the Model 3151 as follows:Function: USER WaveformFrequency: 10.00HzAmplitude: 5VRemote Commands : :RES:OUTP ON:FUNC:MODE USER:TRAC:DEF 4000:TRAC:SEL 1:FREQ:RAST 40e3:VOLT5Using WaveCAD, change the axis (Setup | Axis) to have 4000points (0 to 3999) for the first 4 steps below and 1000 points (0 to999) for step 5 below. Download a 1k or 4k sine waveform (Wave| Sine | OK) into segment 1.FREQ:RAST # of Points Frequency Distortion Reading40e3 4000 10Hz < 0.1%40e4 4000 100Hz < 0.1%40e5 4000 1kHz < 0.1%40e6 4000 10kHz < 0.1%100e6 1000 100kHz < 0.1%Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com