1204. Check the Method of Adjustment: [1 Point Adjustment] or [2 Point Adjustment] .- 1-Point Adjustment: Use when measuring in a range of ± 20.- 2-Point Adjustment: Use when measuring in a wide range.[Initialize] ButtonCheck one[Send Settings]Button5. Enter the actual pre-adjusted reading(s) in the "Current Reading" box and thepost-adjusted reference value(s) in the "Post-adjusted Reading" box.For details see "Rough Guidelines for Adjustment".6. Click the [Send Settings] button to send the adjustment equation to the DataLogger.5. Click on the [Send Settings] Button to send the adjustment equation tothe selected Data Logger.[Initialize] ButtonClick this to clear the adjustment equation from the Remote Unit which is in communicationwith the PC and initialize the unit.Rough Guidelines for AdjustmentNOTE:- The difference between the values entered for "Current Reading" and "Post-adjusted Reading" must be± 5 or less.- Entries are valid to one decimal point.- When using "2-Point Adjustment" if possible select two points apart from one another (we suggest at least10 degrees).- With "2-Point Adjustment" it is not possible to enter values that would result in the slope being 0 (zero).- When using 2-Point Adjustment and measuring in a wide range make sure that the adjustment valuesreflect the wide range being measured.