266 Set up with the display and adjustment module PLICSCOMVEGACAL 65 • Foundation Fieldbus30035-EN-130917Tip:If the display and adjustment module remains on the probe as adisplay, we recommend saving the sensor data in the display andadjustment module.Use the function "Copy sensor data".Start your parameter adjustment with the following menu items of thebasic adjustment:To be on the safe side, note the adjustment values for full and empty.If an adjustment procedure fails, it is not necessary to again carry outa level change.These values can be helpful if the electronics has to be exchanged.% ValueEmpty adjustmentFull adjustmentTab. 1: Adjustment protocolTip:For min. adjustment the vessel should be as empty as possible, andfor max. adjustment, as full as possible. If the vessel is already full,start with max. adjustment.Note:If possible, the vessel should be as empty as possible for min. adjust-ment.Proceed as follows:1. Move from the measured value display to the main menu bypushing [OK].▶ Basic adjustmentDisplayDiagnosticsServiceInfo2. Select the menu item "Basic adjustment" with [->] and confirmwith [OK]. Now the menu item "Min. adjustment" is displayed.Min. adjustment0.00 %=0.0 pF54.5 pF3. Prepare the adjustment value for editing with [OK]. Move to theselection window with [OK].Min. adjustmentAccept current measuredvalue?Accept?Edit?4. Accept the current measured value or move to the editing windowwith "Edit". To edit, set the cursor to the requested position withCarry out min. adjustment