197 Measured value enquiry via Web browser/httpRS232/Ethernet connection • VEGAMET 391/624/625, VEGASCAN 693, PLICSRADIO C6230325-EN-1307015. Enter here an individual password for your connection. This pass-word must be used later on for the dial-up connection.6. Enter the preferred IP address. In general you can use the stand-ard setting "192.168.200.200".7. Enter an individual name in the provided field.Create a dial-up connection on the PCFirst of all create a dial-up connection. Under Windows 2000/Xp thereis the "Assistent for new connection" which you can reach via "Start -Settings - Network connection". Proceed in the way you are creatinga connection to the Internet via modem. Enter under Call numberthe telephone number of the local modem. Enter under User name/Password the specifications already used for the signal conditioninginstrument. Adjust a fix baud rate of 9600.Now start your dial-up connection and establish the communication tothe signal conditioning instrument.Open the web browser (e.g. Internet Explorer) and enter the previ-ously defined IP address in the field called "Address" or "URL". Nowthe HTML chart generated in the signal conditioning instrument isnow displayed with all measured value in your browser window.Fig. 7: Measured value indication by means of Internet Explorer7.5 Measured value enquiry via http softwareTo enquire measured value files, an individual http capable softwarecan be used. With Excel, the measured values can be readout auto-matically e.g. in an individual interval and saved as chart. Different for-mats of the measured value and status files are available. A detaileddescription of these file types is available in chapter "Measured value/Status files".• Signal conditioning instrument with Ethernet interfacePassword/PasswordIP addressHost nameMeasured value indica-tionPrerequisites