Adjustments and Calibrations 6-95. If an adjustment is needed, adjust by the amounts displayed on the front panel.6. After 2 minutes, the test selection menu will re-appear. To examine the testresults after this occurs, highlight the appropriate test and press the Informationbutton. The display is the same as previously except that Do #2: In/Out SkewSetup appears as the first text below the line.When RegiCon #1 or RegiCon #2 FailsThe horizontal alignment error is excessive and a manual adjustment is requiredbefore the diagnostics test routines for fine skew can pass. Perform the CoarseRegiCon Init procedure, see "Coarse RegiCon Initialization" on page 6-12.RegiCon #3 Center Skew AdjustmentThis test adjusts magnification balance. There are no manual adjustments requiredafter completing the test.NoteTo perform this test, move the mark-on-belt sensor to the CENTER position.After the test has been run, the mark-on-belt sensor must be moved back tothe home position.1. Remove the waste cartridge.2. Remove the screw that holds the mark-on-belt (MOB) sensor to the front of theprinter.3. Using the laser lens cleaning tool, move the sensor all the way to the CENTERposition (until it hits the stop).4. Reinstall the waste cartridge.5. Close the door and run the test.6. From the Hidden Service menu, highlight the RegiCon Setup Cycle menuand press OK.7. The first time any RegiCon test is requested, the pages used to perform the testmust be created, so there is a short delay. During this time the front panel displays“Processing Data - Please Wait”8. Highlight Do #3 Center Skew Setup and press OK. The front panel nowdisplays: Testing - Please Wait.