Device data S100 / OPMI picoVersion 9.0Page 248 G-30-1781-enElectromagnetic immunity for all ME equipment and systemsThe S100 / OPMI pico is intended for operation in the electromagneticenvironment specified below. The customer or the user of the deviceS100 / OPMI pico is responsible for ensuring that it is operated in such anenvironment.Electromagnetic immunitytestsIEC 60601 test level Compliance levelElectrostatic discharge(ESD) as per IEC 61000-4-2±8 kVContact discharge±8 kVContact discharge±2 kV, ±4 kV, ±8 kV, ±15 kVContact discharge±2 kV, ±4 kV, ±8 kV, ±15 kVContact dischargeFast transient / burst immunityas per IEC 61000-4-4±2 kV forPower supply lines±2 kV forPower supply lines±1 kV forInput / output lines±1 kV forInput / output linesSurges as perIEC 61000-4-5±0.5 kV, ±1 kVPhase neutral voltage±0.5 kV, ±1 kVPhase neutral voltage±0.5 kV, ±1 kV, ±2 kVPhase / neutral to ground voltage±0.5 kV, ±1 kV, ±2 kVPhase / neutral to ground voltageMagnetic field for supplyfrequency (50/60 Hz) as perIEC 61000-4-830 A/m50 Hz or 60 Hz30 A/m50 Hz or 60 HzVoltage dips, shortinterruptions and voltagevariations as perIEC 61000-4-110 % UT for 1/2 cycleat 0°, 45°, 90°, 135°, 180°, 225°,270° and 315°0 % UT for 1/2 cycleat 0°, 45°, 90°, 135°, 180°, 225°,270° and 315°0 % UT for 1 cycleand70 % UT for 25 cyclesSingle-phase: at 0 degrees0 % UT for 1 cycleand70 % UT for 25 cyclesSingle-phase: at 0 degrees0 % UT for 250 cycles 0 % UT for 250 cycles