Installation163.6 Turn hous ingAfter mounting, the housing can be turned 350° in order to simplify access to the displayand the terminal compartment. Proceed as follows to turn the housing to the requiredposition:• Undo the fixing screws (1)• Turn the housing (2) in the required direction• Tighten up the fixing screws (1). Maximum torque 0.5 Nm.L00-FMU4xxxx-17-00-00-en-0133.7 Ins tallation che ckAfter installing the device, carry out the following checks:• Is the device damaged (visual inspection)?• Does the device correspond to the measuring point specifications for processtemperature, process pressure, ambient temperature, measuring range etc.• If available: Are the measuring point number and labelling correct (visual inspection)?• Is the measuring device sufficiently protected against precipitation and directsunlight?• Are the cable glands tightened correctly?• After aligning the housing, check the process seal at the nozzle or flange.112 2F12 housing T12 housingallen key4 mm/0.1”max. torque 0.5 NmSLP10