CELL-DYN 3000 System Operator’s Manual 6-459140240E — May 1995Chapter 6 Calibration10. When all runs of a sample have been completed, the bulletin linedisplays the message:NEW CAL FACTOR COMPUTED. READY FOR NEXT CALSPECIMENPress [NEXT SPECIMEN] to run the next sample.Calibration Factor CalculationAs each run of a whole blood sample is completed, a Calibration Factoris calculated and updated. This factor is displayed in the column labeled.The column labeled contains the average of thecalibration factors for all whole blood samples used in the calibration.The number in parentheses after mean factor indicates the number ofsamples used to compute that mean factor.The column displays the difference between thecalibration factor currently in the instrument and the mean factorcomputed by the Auto-Cal program.When all runs of all whole blood samples have been completed, themessage:NEW CAL FACTOR COMPUTED – READY FOR ACCEPTANCEis displayed on the bulletin line.1. Press [PREVIOUS SPECIMEN] and/or [NEXT SPECIMEN] toreview and/or print the results of all whole blood samples that wereanalyzed.NOTE: These results are only available for review (andprintout) before the [ACCEPT MEANS] key is pressed.2. Press [PRINT] to obtain a printout of the and FACTORS> and the .3. Enter the factor % difference for each parameter in the CalibrationCriteria Worksheet provided at the end of this procedure. (Thisworksheet may be duplicated as needed.) The Calibration Criteriaare shown in Table 6.4.NOTE: Delete the sign of the factor % difference value beforeentering it on the chart.4. To determine which parameters require calibration, continue withthe next section.