36 USER MANUALTroubleshooting:Display indication Possible cause RemedyInitial heating Tstemperature exceeds105°C, the sensordoes not react whentouched with a fingerThe temperature sensor isdamaged.Contact an authorised servicepoint.Initial heating Tstemperature does notreach 105°C,the halogen heater(s)do not light.The heater is damaged. Replace the heater."Test ..." Auto-tests in progress /electronic unit damagewait for 1 minute" - - - - "The moisture analyzer is duringzeroing/ mechanical damagewait for 1 minutecheck if the moisture analyzer isplaced on stable ground, notaffected by vibrations„Tare range exceeded” Tare key pressed during zeroindicationMoisture analyzer indications mustbe different than zero„Zeroing rangeexceeded”Permissible zeroing range wasexceeded Remove the load from the pan„Weighing rangeexceeded”Permissible weighing range(Max +9e) was exceeded Reduce the load„Measuring rangeexceeded (+)”Upper limit of analog-digitaltransducer measuring range wasexceededRemove the load from the pan„Measuring rangeexceeded (-)”Lower limit of analog-digitaltransducer measuring range wasexceededCheck if there are all necessarypan elements