Action ResultSingle-pulse the remote input. T"TchSPt2 Teaching" displays while thesensor is being taught.TEACH AcceptedThe new value displays on the second lineof the display, flashes, and the sensorreturns to Run mode.TEACH Not Accepted"FAIL" flashes, the sensor returns to step 2,and "RMT TCH" displays.3.6.2 Midpoint TEACHOutput OFF0 mm(500 mm) (700 mm)(600 mm)1000 mmSP1 SP2MidpointOutput ONFigure 15. Window and Midpoint ExampleThe Midpoint TEACH uses both the window size and theTEACH midpoint to determine the actual measurementwindow. For example, a window of 200 mm with a midpoint of600 mm places the measurement window from 500 mm to700 mm.To use Midpoint TEACH:1. Set the window size.2. Set the measurement window usingTEACH Midpointon page 20.The Discrete Output Midpoint TEACH and the Analog OutputMidpoint TEACH are independent settings.Window SizeThe D_OUT > TchMid > WndSize option sets the window size that the Midpoint TEACH uses to set the setpoint one andsetpoint two thresholds.The taught surface must be inside the defined sensing range, and at least one setpoint (with offset applied, if any) must belocated within the sensing range.The Discrete Output window size is a different setting than the Analog Output window size when defined using the pushbuttons.Parameters LTF12 LTF24Window Size Minimum 10 mm 10 mmWindow Size Maximum 11950 mm 23950 mmRange 50 mm to 12000 mm 50 mm to 24000 mmDefault Window Size 20 mm 20 mmNavigate: MENU > D_OUT > TchMid > WndSizeRemote Input: Available1. Access the setup mode.Method Action ResultPush Button Navigate: MENU > D_OUT > TchMid > WndSize . "WndSize" and the current windowsize value display.Remote Inputa. Double-pulse the remote input to entersetup mode.b. Three-pulse the remote input to enterwindow size mode.T TTT TTTTa. "REMOTE SETUP" displays.b. "WndSize" and the current valuedisplay.L-GAGE® LTF Time of Flight Laser Distance Sensorwww.bannerengineering.com - Tel: +1.763.544.3164 19