TroubleshootingDisplay Shows: Cause: Remedy:No segments appear I/O key pressed Press the I/O key.on the display. to turn OFF Display.LL The balance is Check the weighing panunderloaded. for proper installation.Weight is too light or there Increase referenceis no sample on balance. quantity.HH The load exceeds the Unload the balance orbalance capacity. look for obstruction.Display capacity Decrease weightis exceeded. on balance.LLOO Low battery. Replace battery.EE --0022 The calibration operation Place balance in ais being attempted and stable environmentvibration, air drafts or before calibrating.other unstable conditions Before pushing theexist. CAL/MODE button, theLCD should display a stableZERO with the arrow mode.EE --1100 There is a value stored Clear the memory orin the memory when the turn the balance off toTARE button is pressed. restore normal tare function.EE --3300 An electronic malfunction Contact Denver Instrumenthas occurred.EE--5544 This indicates the Contact Denver Instrumentelectronics are no longerwithin the factory setparameters. Objects havedropped onto the weighingpan or the balance itselfhas been dropped. The sensor is out of Contact Denver Instrumentspecification. The balancehas been damaged byrough handling. Liquids orother materials have spilledinto the balance.Do not attempt to access internal components or make any unau-thorized repairs since this will void the warranty.Memo:Note:13