FS10 SeriesFluid Components International LLC 3111TimeDelay toDeactivateRelay orBinary Pulse1000000001- +RELAY_TURN_OFF_DELAY1 – indicates currentvalue relative tomaximum delay(MAX_DELAY)Time delay from when flow measurementis greater/less than relay switch point, toturn off relay. Increments and decrementsin 1 second steps [max. default setting 10seconds when using buttons].*12 AlarmSimulation1100000001- +0000011111= output“max”1111100000= output“min”Alarm simulation mode. Pressing (+)button forces maximum (correspondsto A_OUT_DAC_COUNT_100) 4-20 mAoutput. Pressing (–) button forcesminimum (corresponds to A_OUT_DAC_COUNT_0) 4-20 mA output.Relay output follows the RELAY_POLARsetting.14 FilterSetting1111000001- + INPUT_FILTER_COUNTLED 1 on = setting 3LEDs 1-3 on = 18LEDs 1-5 on = 30LEDs 1-7 on = 50LEDs 1-10 on = 100Input filtering count: Default = 18.(–) decreases filter, min. = 3.(+) increases filter, max. = 100.Filter value 18 reduces responsetime <2 secTable 7 – Button Controls (continued)PC Interface and Command Line Interface ConfigurationsFS10 Communication OptionsIn addition to the button interface, the FS10 may be addressed directly through the RS232 interface. This type of communication offersconvenient access to all user available parameters. There are two methods for interfacing through the RS232 connections:1. A Windows PC Interface program is available that allows reading and writing to all user available functions and parameters.2. A Command Line Interface through a HyperTerminal (MS) type program will access all functions and parameters (see RS232 InterfaceReference Table on the following pages).In both cases the parameters are accessed directly through the RS232 terminals. Depending on the hardware configuration selected, theRS232 signals are available through any of the following:1. M12 pins – for units supplied with M12 power and output (I/O) connector2. 8-wire cable units – for units supplied with cable pigtail, terminate to 3 wires3. 2.5 mm phone jack connector – available on remote mount units onlyNote: RS232 RxD, RS232 TxD and SIGNAL RETURN (3-connections) must be used to for effective communication.An RS232 to 9-pin serial (DB9) to USB interface cable is available from FCI to connect directly to a PC or PLC. An RS232 interface from eitherof the 3 configurations above is available from the table on the following page.Note: The USB interface cable is supplied with a device driver disk. Depending on the computer operating system, installation maybe necessary for communication. Entry of the USB port # will be requested. It is normally the highest value listed.* Maximum time delay is up to 65,000 seconds when using RS232 command line or PC interface.