PCB-BA-e-1315 29AnimalweighingfunctionThe animal weighing function can be applied for busy weighing.During a defined period the average value of the weighing resultsis formed.The more unquiet the weighed item, the longer the period shouldbe selected. In weighing mode keep the PRINT key pressed until [Unit]appears. Press the MODE button several times until „ANL“ is displayed. Acknowledge using SET key, the current setting is displayed. Select the desired settings by pressing the MODE keyANL 3 Period 3 secondsANL 5 Period 5 secondsANL 10 Period 10 secondsANL 15 Period 15 secondsANL off Animal weighing not active Use the SET key to confirm selection. Put the weighing good (animal) on the weighing plate and pressthe SET button. In the display runs a „Countdown“.The average value of the weighing results is displayed andremains displayed on the screen. Use the SET key to change between animal weighing andnormal weighing. Press the SET key to restart the weighing cycle for animalweighing.