• Do not expose the display unit to extreme dampness for longer periods oftime. Non-permitted condensation (condensation of air humidity on theappliance) may occur if a cold appliance is taken to a considerably warmerenvironment. In this case, acclimatize the disconnected appliance for ca. 2hours at room temperature.• Avoid static charge of goods to be weighed or weighing container.Major display deviations (incorrect weighing results) may be experienced shouldelectromagnetic fields (e.g. due to mobile phones or radio equipment), staticelectricity accumulations or instable power supply occur. Change location or removesource of interference.IP 67 protection type as per DIN EN 60529 (KERN UFN only)Suitable for short-term use in wet areas.6.2 UnpackingCarefully remove the display unit from packaging, remove plastic cover andplace it in the designated work area.6.3 Scope of delivery / serial accessories:• Scale• Mains power supply• Operating instructions6.4 PlacingMount the display unit in a way that facilitates operation and where it is easy to see.Accurate weighing results require a weighing bridge with perfect horizontal alignment.During initial installation and after each change of work area it is necessary to levelthe weighing bridge.Use with stand (optional)An optional stand (KERN BFS-07) is available for raising the display unit.UFB/UFN-BA-e-1011 11