7.5 Weighing with tareÖ Deposit weighing vessel. After successful stop check press the button.The zero display and the indicator NET appear.The weight of the container is now internally saved.Ö Weigh the material, the net weight will be indicated.Ö The weight of the weighing container will be displayed as a minus number afterremoving the weighing container.Ö The tare procedure can be repeated as many times as necessary, for examplewith initial weighing of several components for a mix (add-on weighing). The limitis reached when the total weighing range capacity is full.Ö Switch between gross weight and net weight by pressing the key.Ö To delete the tare value, remove load from weighing plate and press .7.6 Weighing with tolerance rangeYou may determine an upper and lower limit for weighing with tolerance limits inorder to ensure that the weighed load remains exactly within the fixed tolerancelimits.During tolerance checks such as dispensing, portioning and sorting, the instrumentwill indicate any lower deviation or exceeding of limits with the help of a visual signalor acoustic signal.Acoustic signal:The acoustic signal depends on the setting of the menu block “BEEP“.Options:• no Acoustic signal turned off• ok Acoustic signal sounds when load is within tolerancelimits• ng Acoustic signal sounds when load is beyond tolerancelimitsUFB/UFN-BA-e-101118