3-8KaVo OP 3D Vision Service Manual032-0358-EN Rev BPrior to beginning the update, ensure that the system is operational and all network connections existand are valid. Previous settings and locations are retained, so it is important to know that the system isfully operational before beginning the update.Follow the installation procedures in the Installation Manual to update clinical software that is requiredat the site. Clinical software should be updated in the following order:• SmartScan STUDIO Integration Services (all sites)• DEXIS i-CAT FLX Server Plugin (DEXIS-FLX sites only)• DEXIS i-CAT FLX Client (DEXIS-FLX sites only)• SmartScan STUDIO Manager (all sites)NOTE: Any non-default permissions set for the SmartScan STUDIO Integration Servicesservices are deleted during an update and will need to be reset.Copy a License File for V-Series Upgradeable Systems1. If license file is on a USB, insert USB into an available USB port on the scanner controller.2. Login to SmartScan STUDIO using the Service Account.3. Select Licenses from the menu.4. For license files on the USB, available license files are displayed. Select license to copy, then press.5. For license files located in a download folder or on a network drive:a. Press .b. Select Download Folder, then select license file. Or, select Network Path, enter the path ofthe license file location and select the license file.c. Press .6. The message License files copied successfully is displayed when complete. Press toreturn to the menu.7. Perform Panel Cal, Shutter Cal, and Geo Cal. Refer to the Installation Manual for instructions.