5-5IEC Command Codes032-0358-EN Rev BX-ray On: Typically equals 1. 0 to (Ticks per frame - 1)X-ray Off: Greater than X-ray On and less than (Ticks per frame - 1)Set as needed for desired length. 50,000 = 1 (ms)Panel Readout: 0 to (Ticks per frame - 1). Typically set the same asX-ray Off, except for continuous X-ray mode.Pulse Mode: 0 or 1. If not entered, defaults to 0.0 = 200 microsecond pulse for panel blanking and valid frame.1 = either a 5msec pulse for valid frame or 3msec for panel blankingTube Range: If not entered, defaults to zero.0 = 120 KV tube, 1 = 90 KV tubeX-Ray CommandsXCN Performs a cine X-ray acquisition for the number of frames requested at the‘Ticks Per Frame’ rate as setup in the X-ray register.Command syntax: XCN Frames WatchdogFrames: 1 to 1024Watchdog: 0 = disabled, 1 = enabledXCT Perform a CT scan for requested distance, speed, and watchdog settings.Command syntax: XCT Distance Speed WatchdogDistance: 1000 to 668563Speed: 1000 to 65000Watchdog: 0 = disabled (default), 1 = enabledXIP Performs a Panoramic scan for the requested parameters.Command syntax: XIP Distance Speed PlatformOffset PlatformStartSpeedPlatformStopSpeed RampReduction WatchdogDistance: 1000 to 668563Speed: 1000 to 70000Platform Offset: 1 to (Distance - 25000)Platform Start: 1000 to 54000Platform End: 1000 to 54000Ramp Reduction: 0 to 250Watchdog: 0 = disabled, 1 = enabledXP Performs an exposed X-ray acquisition for the number of frames requested at the‘Ticks Per Frame’ rate as setup in the X-ray register.Command syntax: XP Frames WatchdogFrames: 1 to 1024Watchdog: 0 = disabled, 1 = enabledMiscellaneous CommandsBD Get Board InformationCommand syntax: BDRetrieves the board component versioning.Results returned in format: OK KRL:x UBT:x FID:x SWB:xKRL: Linux kernel version (short)UBT: U-Boot version (short)FID: 1210 Board Flash (EEPROM) Chip System IDSWB: ROMFS Software Build version (HG-ID Date Time)