MIL-STD-1553/ARINC 429 Triggering and Serial Decode 27Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide 405Searching for MIL-STD-1553 Data in the ListerThe oscilloscope's search capability lets you search for (and mark) certain types ofMIL-STD-1553 data in the Lister. You can use the [Navigate] key and controls tonavigate through the marked rows.1 With MIL-STD-1553 selected as the serial decode mode, press [Search].2 In the Search Menu, press the Search softkey; then, turn the Entry knob to selectthe serial slot (Serial 1 or Serial 2) on which the MIL-STD-1553 signal is beingdecoded.3 Press Search; then, select from these options:• Data Word Start — finds the start of a Data word (at the end of a valid DataSync pulse).• Command/Status Word Start — finds the start of Command/Status word (at theend of a valid C/S Sync pulse).• Remote Terminal Address — finds the Command/Status word whose RTAmatches the specified value. The value is specified in hex.When this option is selected, the RTA softkey becomes available and lets youselect the hex Remote Terminal Address value to find.• Remote Terminal Address + 11 Bits — finds the RTA and the remaining 11 bitsthat match the specified criteria.When this option is selected, these softkeys become available:• The RTA softkey lets you select the hex Remote Terminal Address value.• The Bit Time softkey lets you select the bit time position.• The 0 1 X softkey lets you set the bit time position value as a 1, 0, or X(don't care).• Parity Error — finds (odd) parity bits that are incorrect for the data in the word.• Sync Error — finds invalid Sync pulses.• Manchester Error — finds Manchester encoding errors.For more information on searching data, see "Searching Lister Data" on page 126.For more information on using the [Navigate] key and controls, see "Navigating theTime Base" on page 60.