4-12 Keysight N5290/1A Installation GuidePerformance VerificationSystem Verification4-5. Under Outputs, select the desired output(s). Refer to Figure 4-3.— Print Tabular Data: Prints the verification data in tabular form whichincludes measured data and uncertainty limits. For an example,refer to Figure 4-5 on page 4-14.— Print Graphs: Prints the verification data in graphical form. Thegraphical form includes the measured data trace, factory supplieddata trace, and uncertainty limits. For an example, refer to Figure4-6 on page 4-15.— File Tabular Data: Writes the tabular data to a text file in theWindows 7 directoryD:\SysVer.— File Graphs: Saves a screen image in PNG format in the Windows 7directoryD:\SysVer.To modify the number of ports to be verified or to change thenumber of devices to measure, click on the Configure tab and makethe desired selections.6. Click Run.7. Follow the instructions on the analyzer for performing a full calibration orrecalling an existing recent calibration.8. Follow the instructions on the analyzer for performing the systemverification; inserting the verification devices as prompted.If the System Fails the Verification Test1. Disconnect and clean the device that failed the verification test.2. Reconnect the device making sure that all connections are torqued to theproper specifications.For printed output, it is assumed that the printer has been tested and theWindows driver is installed for the printer that is being used. The systemverification test prints to the printer that has been designated as thedefault printer. (On the Windows Desktop display, click on My Computer,Control Panel, and then Printers to verify the printer setup.)When performing a full calibration, it recommended that the calibration besaved when prompted. If the verification needs to be repeated, thiscalibration can be recalled, saving time.Inspect all connections. Do not remove the cable from the analyzer testport. This will invalidate the calibration that you performed earlier.