Ultimate-cell Stackable Shaking Incubator V2017.2185.11 Extra For Refrigerated UnitsAfter transport or moving –DO NOT USE this unit for at least 24 Hours, unless you are a100% sure that the unit has been moved in an upright position only.If the equipment is continuously used at low temperatures, condensate can occur inside ofthe incubator. It is recommended that you wipe condensations out, or-if possible-heat up theincubator to evaporate this small amount of condensation once every 10 days of operations.To extend the compressors life and to maintain an excellent performance of your refrigera-tion system, the condenser of the unit should be cleaned every month.5.12 Installation of Direct Steam Humidification System (Option)Before starting operation, the direct steam humidification system must be connectedto unpressurized water tank or a pressurized water source with inlet pressure notexceed 0.3 bars.To connect the feed hose to the system, insert the hose into the connector as muchas it goes.5.13 Installation of CO2 gas supply hose (Option)The inlet nozzle for CO2 gas supply is on the right side of the case wall. To connectthe CO2 gas supply hose to the nozzle, make sure: Inlet pressure not exceed 3.0 Bar. (2.0 Bar recommended) Gas supplied must be dry, clean, dust and oil free. The hose must remain intact and pressure resistant. Keep the hose secured by a hose clamp.ATTENTION:The nominal cooling capacity can only be reached if the ambient tempera-ture at the cooling aggregate is below 30 °C.Clean the shaking incubator before you put it into use and on a regular base.The interior should be wiped down with an appropriate disinfectant, suchas 70 % ISOPROPYL ALCOHOL or equivalent.DO NOT USE ANY CHLORINATED OR HALOGEN MATERIAL-AS THIS IS HARMFUL TO THE POLISHED STAINLESS STEEL!!!ATTENTION!Use only distilled or deionized water. Using water outside this range willvoid your warranty.