ECO GOLD28.04.10/ YACE0088 Putting the device into operation 517.7.2 Setpoint offset operating mode (Diff.set/actual)It is possible to apply an offset value to the temperature, which is provided by an external temperaturesensor and to process it as the set value. The bath temperature can therefore be operated, for example,-15 °C below the temperature of a reactor measured by the external temperature sensor.− Access to the menu level is obtained by pressing .− Selection and confirmation of Æ Setup Æ Control ÆSetpoint offset. Adjacent menu window appears.− As well as Diff.set/actual value the currently set offset valueis displayed.− The adjacent menu window appears on selecting the menupoint "Offset source".− Select the set-value source with or and confirm itwith .− The setpoint offset is deactivated with "off".− Select Diff.set/actual value with or and confirmwith . The adjacent menu window appears.− The minimum and maximum possible offset values and thecurrent offset value are displayed.− Change the value with or .− By pressing or you can also select numbers indi-vidually and change them with or .− By pressing (+/-) the arithmetic sign "+" or "-" can beselected.− You confirm the set value by pressing .− You quit the window without changes with (ESC).ESC ּס OK StandbySetpoint offsetOffset source ►Diff.set/actual value 0.0ESC ּס OK +/-Offset valueDiff.set/actual valueMax:500,0Min:-500,00,0ESC ּס OK StandbyOffset sourceoffextern Pt100