ECO GOLD92 Technical data and graphs 28.04.10/ YACE008811 Technical data and graphsThe figures were determined according to DIN 12876.Data applicable to all ECO thermostatsAmbient temperature range °C 5 ... 40Setting resolution °C 0.01Display resolution °C 0.01Absolute accuracy K ±0,2Relative humidity maximum relative humidity 80 % at 31 °C and decreasing linearly to50 % up to 40 °CTemperature for storage °C -20…45 (for air-cooled thermostats)Pump type/number of powerlevels Pressure pump/6Discharge pressure, max. bar 0.55Discharge flow, max. L/min 22Heatingpower230 V; 50/60 Hz kW 2.6Power con-sumption 230 V; 50/60 Hz kW 2.7Class of protection IP 21Safety equipment Class III/FL to DIN 12876-1; suitable for flammable and non-flammable liquidsClass of protection for electricaloperating equipment DIN EN61140 (VDE 0140-1)Class IEMC requirements to DIN EN61326-1 (corresponds to VDE0843-20-1)applies to EuropeClass Bfor Canada and the USA Class AEC directives The devices conform to the directives of the European Parliamentand of the Council: 2004/108/EC regarding electromagnetic com-patibility (EMC) and 2006/95/EC relating to electrical operatingequipment for use within certain voltage limits (Low Voltage Direc-tive).The devices bear the CE label.