ECO GOLD52 Putting the device into operation 28.04.10/ YACE00887.8 ProgrammerThe programming function enables you to save five temperature/time programs. The programs consist ofseveral temperature/time segments. This also includes how often the program is to be run (loops). Thesum of all segments of all programs can be up to 150. Step changes in temperature, i.e. the time is zero,or also temperature holding phases, i.e. the temperature at the start and end of a segment is the same,are possible. On starting the current set value is taken as the starting value of the first segment.If the pump level is to be changed in the course of the program, then the specified pump level is enteredin the respective program line. If the pump level is to remain unchanged, "0" is entered and it is indicatedas "---" in the display.7.8.1 Programming exampleNo. Tend hh mm Tolerance No. Pump S1 S2 S3Start 30.00 -- -- 0.1 Start 2 off off off2 50.00 0 20 0.0 2 2 off off off3 70.00 0 40 0.0 3 3 off off off4 70.00 0 10 0.1 4 4 off off off5 60.00 0 30 0.0 5 2 off off off6 30.00 0 0 0.0 6 2 off off off No time specification is possible in segment 1 (starting segment). The temperature of the firstsegment is attained as quickly as possible in order to switch to segment 2 after reaching theset tolerance.Cooling time dependent on de-vice type, load etc.Graph 1